| Output signal | 4-20mA DC signal linear with temperature within the range |
| Basic accuracy | Thermocouple temperature transmitter±0.5%FS; Thermal resistance temperature transmitter ±0.2%FS; |
| Transmission mode | two-wire system |
| DC power supply | 12-30V DC. rated voltage 24V DC |
| Load impedance | Limit load of load resistance calculated; RL(max)=50(U-12) In the formula, the transmitter is a DC power voltage Under 24V rated voltage: RL(max)=50(24-12)=600Ω |
| Normal operating environments | a.Ambient temperature: - 25℃~85℃ ( special requirement - 40℃~90℃ ) b. Relative humidity: 5%-95% c.Mechanical vibration: 10-150HZ. A=0.15mm d. No medium caused corrosion to the transmitter in the ambient air. |
| Electromagnet interference | When the radio transmitter of 20-1000MHz and 5W 0.5 meters away from the transmitter works.lt shall vary ± 0.15%FS. |
| Temperature drift | When the environment temperature varies by 1℃, the zero point varies ±0.15% FS and the measuring rang changes ± 0.01 %FS |
Use and wiring
1. Notice to wiring: Please note that the temperature-sensing element has been well connected and the measurement range before leaving the factory, Generally, please do not adjust the range adjustment potentiometer, unless it is necessary.
2. For the installation of integrated TC and RTD, please refer to other operation instructions for the same types of TC and RTD.
3. Integrated TC and RTD are provided with a two-wire system current output, so only two copper conductors shall be required.
Example for order
Please indicate the model, graduation mark, total length and inserted depth in order,i.e.(L×I)
Example1:WRNKB-230 K 1150x1000
I.E: Two wire system integrated sheathed thermocouple, fixed bolt installation, waterproof connection box, outer diameter Φ16mm, graduation mark K, overall length 1150, inserted depth 1000mm.
Example 2: WZPKB-336 Pt100 L=1000
I.E: Two-wire system integrated sheathed platinum thermal resistance, movable ferrule bolt installation, waterproof connection box, outer diameter, graduation mark Pt100, overall length 1000mm.
Thermocouple and thermal resistance with temperature transmitter
| Model | Graduation Mark | Measuring Range(℃) | Casing tube | ||
| Material | Tube dia. | Spec. | |||
| SBWZPR-231 | Pt100 | -50-+400 | 1Cr18Ni9Ti、316 | Φ6-12 | 300×150 400×250 450×300 500×350 |
| SBWRNR-231 | R | 0-1000 | 1Cr18NiSTi、GH2520、316 | Φ6-12 | |

| Model | Graduation Mark | Measuring Range(℃) | Casing tube | ||
| Matenial | Tube dia. | Spec. | |||
| SBWZPK2-231 | Pt100 | -50-+400 | 1Cr18Ni9Ti、316 | Φ6-12 | 550×400 600×450 650×500 700×550 750×600 800×650 1000×850 1150×1000 1350×1200 |
| SBWRNR2-231 | R | 0-1000 | 1Cr18Ni9Ti、GH2520、316 | Φ6-12 | |

Hirschmann plug-in type small thermal resistance and thermocouple
| Model | Graduation Mark | Measuring Range(℃) | Casing tube | ||
| Matenial | Tube dia. | Spec. | |||
| SBWET-201 | Pt100 | -50-+400 | 1Cr18Ni9Ti、316 | Φ4-12 | 50 100 150 200 300 400 500 |
| SBWET-200 | R | 0-800 | 1Cr18Ni9Ti、GH2520、316 | Φ4-12 | |

As China Integrated temperature transmitter Suppliers and Integrated temperature transmitter Factory, Jiangsu Zhaolong Electrics Co., Ltd. is a regional high-tech enterprise and a production base of temperature sensors and temperature control instruments. We have jointly developed and produced a series of thermocouples (thermal resistances) and temperature control electronics with numerous R&D institutions and academies in Shanghai. These products are salable throughout the country and the world, offering an advantage in cost performance. Our products have a renowned reputation in the market and win customers' trust due to their excellent performance, attractive appearance, high reliability, advanced technology, and good cost performance.
With strong technical development team and patent results, we can undertake business for normal and special specifications, and can design & install various large-scale and small-scale automatic temperature control boxes and temperature control rooms and remote monitoring temperature control systems for customers.
An ambient air temperature sensor measures the temperature of the surrounding air at a specific location and converts that measurement into an electrical signa...
READ MOREA pyrometer measures temperature—specifically the temperature of objects and surfaces without requiring any physical contact. Unlike conventional thermometers ...
READ MOREHow to test a pressure transducer in real engineering environments typically follows a structured path: applying a standard pressure input, comparing output sig...
READ MOREHow to calibrate temperature sensor typically relies on comparing the measured value with a standard temperature reference and correcting the deviation to impro...
READ MOREHow to calibrate temperature sensor typically involves using a stable and known temperature source, such as a temperature bath, dry block calibrator, or ice poi...
READ MORE